ProductGuru Recalls

Official recall record

REMISOL Advance UDI-DI code: 13700962601874 All Software Versions: B92487; B92488; C24317; C28652; C37500; D04164; C44703; C57017; C69412

US FDA — Medical Devices · 2025-04-08 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-04-08
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
13700962601874
Brand on record
Beckman Coulter Inc.
Product category
Medical Device

What the authority reported

Due to a software issue, the software may incorrectly process the date of birth (DoB) received from the Laboratory Information System (LIS) and when the DoB is set to 01/01/1900 or left empty, REMISOL Advance may incorrectly identify a patient as a newborn, affecting the execution of validation rules based on incorrect reference ranges. The test results are not changed by this software defect and are accurately displayed by REMISOL Advance; however, the evaluation flagging of the results may be affected. This issue may lead to wrong reference range calculations and can result in erroneous flagged results being auto validated and uploaded to LIS.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.