ProductGuru Recalls

Official recall record

UIBC (Unsaturated Iron Binding Capacity), REF: OSR61205,

US FDA — Medical Devices · 2025-06-26 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-06-26
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
15099590011925
Brand on record
Beckman Coulter Inc.
Product category
Medical Device

What the authority reported

Beckman Coulter identified that the Unsaturated Iron Binding Capacity (UIBC) assay is not meeting their labeled hemoglobin interference claims of no significant interference (i.e., less than 10%) up to 200 mg/dL hemoglobin as indicated in their instructions for use. Initial internal testing confirmed that samples with low UIBC levels failed the claimed hemolysis interference specifications with reported biases up to -43.6 % UIBC when hemolyzed serum samples contained 200 mg/dL hemoglobin.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Same barcode in other registries

The same EAN appears in 1 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.

Other recalls from this brand

All 32 official recall notices naming Beckman Coulter, Inc. on this site →

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.