CASE v7.0 Cardiac Testing System Model no. 3031234-001 Product Code SU3

US FDA — Medical Devices · 2026-04-24 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2026-04-24
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00195278617644
Brand on record
GE Medical Systems Information Technologies Inc

What the authority reported

Electrical sparking may occur at the printer head. If the Electrostatic Discharge (ESD) printer brush becomes dislodged, electrical sparking at the printer head can result in charring or discoloration of thermal printer paper. Ignition of the paper could potentially result in serious thermal injury.