BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, for FILMARRAY Systems, when used in conjunction with Culture Media

US FDA — Medical Devices · 2024-06-19 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-06-19
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00815381020338
Brand on record
BioFire Diagnostics, LLC

What the authority reported

If blood culture identification panel is used in conjunction with specific lots of culture media bottles, then false positive Serratia marcescens results may occur, due to an increased level of non-viable organism from serratia marcescens targets in culture media bottles, false positive result may lead to an inappropriate change in patient therapy.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Same barcode in other registries

The same EAN appears in 2 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.

Other recalls from this brand

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.