Official recall record
BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, for FILMARRAY Systems, when used in conjunction with Culture Media
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-06-19
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00815381020338
- Brand on record
- BioFire Diagnostics, LLC
What the authority reported
If blood culture identification panel is used in conjunction with specific lots of culture media bottles, then false positive Serratia marcescens results may occur, due to an increased level of non-viable organism from serratia marcescens targets in culture media bottles, false positive result may lead to an inappropriate change in patient therapy.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Same barcode in other registries
The same EAN appears in 2 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2023-10-17 · US FDA — Medical Devices
BIOFIRE Blood Culture Identification 2 (BCID2) Panel, REF: RFIT-ASY-0147, used with Maine Molecular Quality Controls, In - 2021-10-11 · US FDA — Medical Devices
BioFire Blood Culture Identification 2 (BCID2) Panel Part No: RFIT-ASY-0147 UDI: 00815381020338 Blood Culture Catalog N
Other recalls from this brand
- 2024-08-07 · US FDA — Medical Devices
FILMARRAY 2.0 INSTRUMENT, REF: FLM2-ASY-0001, FLM2-ASY-0001-W