ProductGuru Recalls

Official recall record

Calibrate (80 kVp) nanoDot D2DNS, Model Numbers: a) 18121-000; b)18121-KIT; c) 18121-SET; radiation monitoring dosimeter used with th

US FDA — Medical Devices · 2023-07-12 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2023-07-12
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
0860003399903
Brand on record
Landauer
Product category
Medical Device

What the authority reported

LANDAUER received reports indicating that some nanoDots may potentially be outside of the specified range of +/-5.5% accuracy. Investigation of nanoDots indicated the presence of a potential non-conformance in the Optical Stimulated Luminescence (OSL) material that is beamed after exposure and emits fluorescence proportional to radiation exposure.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Same barcode in other registries

The same EAN appears in 6 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.

Other recalls from this brand

All 28 official recall notices naming Landauer on this site →