Official recall record
MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2026-05-27
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00763000521554
- Brand on record
- Medtronic MiniMed, Inc.
What the authority reported
A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Verify at the source:
US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-09-16.
Record last checked against the registry on 2026-09-16.
Other recalls from this brand
- 2026-08-20 · US FDA — Medical Devices
Brand Name: MiniMed Flex Product Name: MiniMed Flex Insulin Pump Model/Catalog Numbers: MMT-8062; MMT-8072 Pro - 2026-07-13 · US FDA — Medical Devices
MiniMed 640G Insulin Pump, Model: MMT-1711, MMT-1712; - 2026-07-13 · US FDA — Medical Devices
MiniMed 670G Insulin Pump, Model: MMT-1781, MMT-1782;
All 30 official recall notices naming Medtronic MiniMed, Inc. on this site →
ProductGuru Recalls republishes official safety information with attribution. We are not
the issuing authority and this page is not advice. If this product is a medical device or
medication, speak to a doctor or pharmacist. Found an error?
Report it — corrections are published.