ProductGuru Recalls

Official recall record

MiniMed 780G Insulin Pump, REF: MMT-1885, MMT-1886, MMT-1895, MMT-1896

US FDA — Medical Devices · 2026-05-27 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2026-05-27
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
00763000521554
Brand on record
Medtronic MiniMed, Inc.

What the authority reported

A software anomaly can occur when the pump is paired with the Instinct sensor. Under certain circumstances, the pump unexpectedly stops delivering insulin and displays two error messages back-to-back: Pump Error 53 followed immediately by Pump Error 23. Insulin delivery will stop which may lead to delayed therapy and an increased risk of hyperglycemia or diabetic ketoacidosis (DKA).

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-09-16.