ProductGuru Recalls

Official recall record

Sevoflurane E_Vap.(Quik-Fil):used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentra

US FDA — Medical Devices · 2024-07-22 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-07-22
Country of alert
United States
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
06936415975784
Brand on record
Mindray DS USA, Inc. dba Mindray North America

What the authority reported

V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters

Other recalls from this brand

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.