Official recall record
Desflurane E_Vap.(Saf-T-Fill) used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic agent with controllable concentr
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-07-22
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 06936415975760
- Brand on record
- Mindray DS USA, Inc. dba Mindray North America
What the authority reported
V90 electronic vaporizer anesthetic agent output may be higher than the set amount due to a component within the vaporizer causing an incomplete seal in some cases. Should this issue occur, the A9 Anesthesia System will generate alarms associated with abnormal vaporizer output. If the anesthetic concentration is too high, the patient can inhale too much anesthetic agent; the patient may suffer from symptoms such as vertigo and drowsiness after the operation. In severe cases, the patient may suffer from neurological sequelae such as hypomnesia and inattention
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2024-07-22 · US FDA — Medical Devices
Sevoflurane E_Vap.(Quik-Fil):used with the A9 Anesthesia System-Used to provide the vapor of an anesthetic age - 2024-07-22 · US FDA — Medical Devices
Sevoflurane E_Vap.(Safety Filling adapter) used with the A9 Anesthesia System-Used to provide the vapor of an - 2024-07-22 · US FDA — Medical Devices
Isoflurane E_Vap.(Safety Filling adapter): used with the A9 Anesthesia System-Used to provide the vapor of an