Power Express, REF B90918

US FDA — Medical Devices · 2024-07-17 · GermanyClass II
Issuing authority
US FDA — Medical Devices
Alert date
2024-07-17
Country of alert
Germany
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
15099590681555
Brand on record
Beckman Coulter Biomedical GmbH

What the authority reported

The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.