PE Centrifuge Temperature Controlled, REF: B36366, a component of the Power Express.
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2024-07-17
- Country of alert
- Germany
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 15099590673567
- Brand on record
- Beckman Coulter Biomedical GmbH
What the authority reported
The adhesive joint between the plastic socket and metal component of the Power Express Sample Holder can fail, causing the metal part to detach and enter the centrifuge. This detachment can result in damage to the centrifuge and potential destruction of the patient sample resulting in delayed results and possible exposure of lab technician to biohazardous material.