DxC 700 AU, REF: B86444, B86446

US FDA — Medical Devices · 2026-01-30 · JapanClass II
Issuing authority
US FDA — Medical Devices
Alert date
2026-01-30
Country of alert
Japan
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
14987666542545
Brand on record
Beckman Coulter Mishima K.K.
Product category
Medical Device

What the authority reported

A delay in results may occur. When clinical chemistry analyzer calibration monitor with PC installed with Microsoft SQL Server 2022 Express is open beyond15 minutes may lead to: Reagent Blank and calibration databases not opening errors: CAL History DB Open Error.., RB History DB Open Error..; system will stop analyzing, firm service must visit to apply software solution; may lead to delay in diagnosis and/or patient treatment.