Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 - 17943367 (ref Z-1907-2025)
- Issuing authority
- US FDA — Medical Devices
- Alert date
- 2025-04-30
- Country of alert
- Japan
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 14987666540664
- Brand on record
- Beckman Coulter Mishima K.K.
What the authority reported
A manufacturing defect in the sample probe (S probe). The corners of the tip of the defective S probe are rounded and chamfered, which can prevent the proper formation of the "air layer" required for normal dispensing. The air layer is the air space that separates the cleaning water filled in the probe from the sample sucked into the probe. Maintaining this air space prevents the cleaning water and the sample from mixing. In the defective S probe, the "air layer" was not properly formed, causing the cleaning water and the sample to mix diluting the sample. As a result, the sample mixture dispensed into the cuvette has a lower concentration than the original sample, and the analyte concentration measured is lower than expected results.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2026-08-03 · US FDA — Medical Devices
Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: - 2026-01-30 · US FDA — Medical Devices
CHEMISTRY ANALYZER AU5800, REF: B96697, B96698, B23279, B23280, B23281 - 2026-01-30 · US FDA — Medical Devices
DxC 700 AU, REF: B86444, B86446