ProductGuru Recalls

Official recall record

Systems/units containing affected software/SQL BECKMAN COULTER CHEMISTRY ANALYZER AU480 REF / UDI-DI code: B96692 / 14987666542965 B9669

US FDA — Medical Devices · 2026-08-03 · JapanClass II
Issuing authority
US FDA — Medical Devices
Alert date
2026-08-03
Country of alert
Japan
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
14987666542965
Brand on record
Beckman Coulter Mishima K.K.
Product category
Medical Device

What the authority reported

Certain AU480 chemistry analyzers, with Windows 10 IoT Enterprise 2021 LTSC PCs, may output the message System Stopped By Fatal Error: DBConnectionError AlarmDB when launching the instrument software, using PC pre-installed with Microsoft SQL Server 2022 Express. When this occurs, AlarmDB becomes unreadable and the instrument software cannot be started. This causes a delay in obtaining patient results.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-10-22.

More recalls: US FDA — Medical Devices · Japan · all filters

Other recalls from this brand