iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-08-13
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Brand on record
- Beta Bionics, Inc.
What the authority reported
it was found that in versions 1.3.7, 1.4.2, and 1.4.3, the Lock Screen and Limited Access Passcode Screen on the iLet graphical user interface (GUI) include certain icons displayed in the status bar that are active, thereby allowing the user to bypass those screens when those icons on the status bar are pressed, allowing unauthorized access while the device is in Limited Access Mode. A Health Risk associated with Limited access mode includes severe hypoglycemia due to unauthorized access to the iLet if someone were to make unauthorized meal announcements or stopped insulin delivery.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Other recalls from this brand
- 2026-03-12 · US FDA — Medical Devices
iLet Bionic Pancreas, REF: BB1001 - 2025-09-23 · US FDA — Medical Devices