Official recall record
iLet Bionic Pancreas, REF: BB1001 (ref Z-2513-2026)
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-09-23
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Brand on record
- Beta Bionics, Inc.
What the authority reported
The reason for the recall is: it was found in version 3.3.6 of the iLet ACE Pump firmware and all previous releases that a screen unlock failure could occur when the user successfully slides the unlock slider on the Lock Screen, no screen transition occurs, and the user remains on the Lock Screen. Because the user cannot exit the Lock Screen, the user is unable to perform a cartridge change or meal announcement. If this occurs and therapy is in progress, therapy will continue normally until the insulin cartridge is depleted. The user will be unable to perform a cartridge change or a meal announcement and will be stuck on the screen until the battery fully drains. This can lead to hyperglycemia.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-09-16.
Other recalls from this brand
- 2026-03-12 · US FDA — Medical Devices
iLet Bionic Pancreas, REF: BB1001 - 2025-08-13 · US FDA — Medical Devices
iLet Ace Pump Kit REF: BB1001 iLet Dosing Decision Software