ProductGuru Recalls

Official recall record

iLet Bionic Pancreas, REF: BB1001

US FDA — Medical Devices · 2026-03-12 · United StatesClass II
Issuing authority
US FDA — Medical Devices
Alert date
2026-03-12
Country of alert
United States
Alert level
Class II
Alert type
recall
Brand on record
Beta Bionics, Inc.

What the authority reported

insulin pump software versions 1.4.2, and 1.4.3 issue with Dexcom Continuous Glucose Monitoring (CGM) G7 sensor may result in delayed CGM reading then switch to non-CGM dosing, leading to only basal/requested meal doses provided, no correction doses/insulin reductions/suspensions for high/low glucose unless blood glucose entered for sensor life or replacement. Potential health risks to users include severe hyperglycemia with and without DKA and severe hypoglycemia.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · United States · all filters