Sample probe sucks a sample dispensed into a tube or cup and discharges the appropriate amount into a cuvette. Lot 178713114 - 17943367 (ref Z-1908-2025)

US FDA — Medical Devices · 2025-04-30 · JapanClass II
Issuing authority
US FDA — Medical Devices
Recall initiated by firm
2025-04-30
Country of alert
Japan
Alert level
Class II
Alert type
recall
Barcode (EAN/GTIN)
14987666540732
Brand on record
Beckman Coulter Mishima K.K.

What the authority reported

Sample probes with defective probe tips, when used with clinical chemistry analyzers, may lead to a lower concentration being dispensed, which may not be detectable by QC checks, which could lead to the reporting of false low test results (Max 34% error at 1.0 microliters and 7% error at 1.6 microliters sample dispensing volume), which may lead to delayed recognition and treatment.

What to do

Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.

Verify at the source: US FDA — Medical Devices — official record →
Record last checked against the registry on 2026-05-31.

More recalls: US FDA — Medical Devices · Japan · all filters

Other recalls from this brand

All 7 official recall notices naming Beckman Coulter Mishima K.K. on this site →

ProductGuru Recalls republishes official safety information with attribution. We are not the issuing authority and this page is not advice. If this product is a medical device or medication, speak to a doctor or pharmacist. Found an error? Report it — corrections are published.