Official recall record
Medtronic SmartLink software loaded on CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1.
- Issuing authority
- US FDA — Medical Devices
- Recall initiated by firm
- 2025-06-17
- Country of alert
- United States
- Alert level
- Class II
- Alert type
- recall
- Barcode (EAN/GTIN)
- 00763000002039
- Brand on record
- Medtronic, Inc.
What the authority reported
A software update for Medtronic CareLink SmartSync Device Manager (SmartSync) is now available. The update eliminates the potential display of an erroneous electrical reset (pop-up) message upon interrogation with a SmartSync Device Manager. The erroneous message indicates an electrical reset has occurred. However, no reset has actually occurred and there are no parameter changes or loss of diagnostics. Prior to this software update, if an erroneous SmartSync message was displayed, confusion regarding the message could have caused a clinician to consider an unnecessary system revision. Medtronic has received 39 reports of this display error with two instances of unnecessary device explant through 30 April 2025.
What to do
Follow the instructions in the official notice — recall remedies (refund, repair, replacement, or disposal) are defined by the issuing authority and the responsible business, and they differ per recall. Do not rely on summaries, including this one.
Record last checked against the registry on 2026-05-31.
Same barcode in other registries
The same EAN appears in 3 other official recall record(s) — a cross-border signal worth checking if you bought this product abroad.
- 2025-06-18 · US FDA — Medical Devices
Medtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components: 1. Cobalt - 2025-06-17 · US FDA — Medical Devices
Medtronic SmartLink Software loaded on SmartSync Patient Connector, Model Number 24967 with the following software compo - 2025-06-18 · US FDA — Medical Devices
Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1. Cobalt Cr
Other recalls from this brand
- 2026-03-11 · US FDA — Medical Devices
Medtronic Sphere-9 Catheter, Model Number AFR-00001; catheter, percutaneous, cardiac ablation, for treatment o - 2025-06-18 · US FDA — Medical Devices
Medtronic CareLink SmartSync Patient Connector, Model Number 24967 with the following software components: - 2025-06-18 · US FDA — Medical Devices
Medtronic CareLink SmartSync Device Manager, Model Number 24967A with the following software components: 1.
All 6 official recall notices naming Medtronic, Inc. on this site →